NJ ATM 1 - Particulate Emissions




Method Overview

  • Isokinetic.
  • Heated out of stack filter (M5).
  • "Condensible" particulates pass through the filter and condense in impingers. Can readily be combined with a CPM Method.
  • Tester may adjust out-of-stack filter temperature to suit source specific conditions.

Target Compounds

  • Particulate Matter (PM) -- those substances retained in the sampling nozzle/probe and on a heated filter after the removal of uncombined water.

Performance Tips

  • A minimum sample catch of 50 mg is targeted to calculate the minimum sample times
    • An absolute minimum of 3 mg catch is required;
    • Any expected catch of less than 50 mg will require the use of EPA Alternative Method 005 to process the samples.
  • During sampling a filter is maintained at a specified temperature (248oF + 25oF)
  • Do not use silicone grease - it can be soluble in the probe rinse and can bias the PM value high.
  • Run the train isokinetically.
  • Stainless steel nozzles should be used with great care, port scrapings can bias results high.

Shipping

  • Acetone must be shipped as a hazardous material. Please call for IATA codes if you need them.
  • Samples may be shipped at ambient temperatures.
  • It may prove more economical to keep acetone rinses separate, ship other samples normally, and the acetone as hazardous.
  • Be sure to include a chain of custody for each shipping container.